EquipmentLaboratoryBinder Lab Incubator BD240

Binder Lab Incubator BD240



STOCK REF#: G249


Binder BD240 Incubator

Drying ovens/Heating ovens

Heating oven with mechanical control

Incubator with gravity convection

The BINDER incubator of the BD series with gravity convection is the specialist in
long-term and stable continuous operation. This incubator is specially designed
for gentle incubation of organisms, as well as conditioning of heat-sensitive
media or microbiological heating.

Advantages:

    Safe and reproducible incubation
    Disinfection routine
    Broad range of products and applications

Electronically controlled APT.line™ preheating chamber assuring temperature
accuracy and reproducible results

Temperature range from 5 °C above room temperature to 100 °C

DS control with integrated timer 0 to 99 hrs

Digital temperature setting with an accuracy of a tenth of a degree

Independent temperature safety device class 3.1 (DIN 12880) with optical alarm

Adjustable ventilation by means of front ventilation flap slide and rear exhaust Ø
50 mm

Inner glass doors

RS 422 interface for communication software APT-COM™ DataControlSystem

2 chrome-plated racks included

Exterior dimensions
   
    Housing:     
    Width (mm) 1040 Height (incl. feet/castors) (mm) 880
    Depth (plus connections and fittings) (mm)     750
    Depth (incl. connections and fittings) (mm)     840
    Wall clearance, rear (mm)     100
    Wall clearance, side (mm)     160
    Steam space volume (l)     283
    Number of doors (ea.)     2
            
    Interior dimensions     
    Width (mm) 800     Height (mm) 600 Depth (mm) 500
    Interior volume (l)     240
    Racks (number standard/max.)     2/7
    Load per rack (kg)     30
    Permitted total load (kg) 70
    Weight (empty) (kg)     93
            
    Temperature data     
    Temperature range from 5 °C above ambient temperature to 100°C
    Temperature variation     
    at 37 °C (± K)     0,5
    at 50 °C (± K)     0,9
    Temperature fluctuation     
    at 37 °C (± K)     0,1
    at 50 °C (± K)     0,1
    Warm-up time 1)     
    to 37 °C (min.)     70
    to 50 °C (min.)     115
    Recovery time after doors were open for 30 sec. 1)     
    at 37 °C (min.)     5
    at 50 °C (min.)     6
            
    Electrical data     
    IP protection class acc. to EN 60529     IP 20
    Voltage (± 10%) 50 / 60 Hz (V)     230
    Nominal power (kW)     0,68
    Energy consumption at 37 °C (W)     33

1) To 98% of the set value

Manufacturer: Binder

POA - Contact Us



The pictures you are looking at are of the Binder Lab Incubator BD240. This is the actual item you are buying, they are not downloaded or copied from any other source.
All information given on this website is correct at the time of publication. However in the interests of technical progress, design specifications are subject to change without notice. Accordingly, data given herein should be regarded as a general guide only and does not form part of any contract. All goods are supplied in accordance with our standard terms and conditions of sale. If you need more information on using this equipment check out The Laboratory Encyclopedia.
Back Top
Services | Shipping | About | FAQ | Terms & Conditions | Testimonials | Links | Contact Us |
Content

LoginSafe to pay online

Or use an external account